Pulsed direct and constant direct currents in the pilocarpine iontophoresis sweat chloride test

BMC Pulm Med. 2014 Dec 13:14:198. doi: 10.1186/1471-2466-14-198.

Abstract

Background: The classic sweat test (CST) is the golden standard for cystic fibrosis (CF) diagnosis. Then, our aim was compare the production and volume of sweat, and side effects caused by pulsed direct current (PDC) and constant direct current (CDC). To determine the optimal stimulation time (ST) for the sweat collection. To verify the PDC as CF diagnosis option.

Methods: Prospective study with cross-sectional experimental intervention. Experiment 1 (right arm): PDC and CDC. ST at 10 min and sweat collected at 30 min. Currents of 0.5; 0.75; 1.0 and 1.5 mA and frequencies of 0, 200, 1,000 and 5,000 Hz applied. Experiment 2 (left arm): current of 1.0 mA, ST at 5 and 10 min and sweat collected at 15 and 30 min with frequencies of 0; 200; 1,000 and 5,000 Hz applied Experiments 1 and 2 were performed with current density (CD) from 0.07 to 0.21 mA/cm2. Experiment 3: PDC was used in typical CF patients with two CFTR mutations screened and or with CF diagnosis by rectal biopsy and patients with atypical CF.

Results: 48 subjects (79.16% female) with average of 29.54 ± 8.87 years old were enrolled. There was no statistical difference between the interaction of frequency and current in the sweat weight (p = 0.7488). Individually, positive association was achieved between weight sweat and stimulation frequency (p = 0.0088); and current (p = 0.0025). The sweat production was higher for 10 min of stimulation (p = 0.0023). The sweat collection was better for 30 min (p = 0.0019). The skin impedance was not influenced by ST and sweat collection (p > 0.05). The current frequency was inversely associated with the skin impedance (p < 0.0001). The skin temperature measured before stimulation was higher than after (p < 0.0001). In Experiment 3 (29 subjects) the PDC showed better kappa index compared to CDC (0.9218 versus 0.5205, respectively).

Conclusions: The performance of the CST with CDC and PDC with CD of 0.14 to 0.21 mA/cm2 showed efficacy in steps of stimulation and collection of sweat, without side effects. The optimal stimulation time and sweat collection were, respectively, 10 and 30 min.

Publication types

  • Comparative Study

MeSH terms

  • Adult
  • Biopsy
  • Chlorides / analysis
  • Cystic Fibrosis / diagnosis*
  • Electric Impedance
  • Electrodiagnosis / adverse effects
  • Electrodiagnosis / methods*
  • Female
  • Humans
  • Iontophoresis / adverse effects
  • Iontophoresis / methods*
  • Male
  • Muscarinic Agonists / pharmacology
  • Pilocarpine / pharmacology
  • Prospective Studies
  • Rectum / pathology
  • Sweat / chemistry
  • Sweat / metabolism*
  • Sweat Glands / drug effects
  • Sweat Glands / physiopathology
  • Time Factors
  • Young Adult

Substances

  • Chlorides
  • Muscarinic Agonists
  • Pilocarpine